Home > News > Equipment and instruments >

How to set the temperature of the GX series drying oven?

In order to reduce everyone's confusion, Kenton will take GX series as an example to explain to you how to set the temperature of the drying oven?

The first thing to look at is the meaning of the parameters of the control panel, the following image.

                    \
                                             \

These buttons, with reference to the text above, memorize the definition of each button.

1. “start” button: start or stop

2. “set” button: set or view temperature, time and other settings.

3. “shift/self regulating” button: in non-setting mode, long press this button for 6 sec. to enter or exit system auto-setting; in setting mode, press this button to modify set value.

4. “decrease/rerun” button: in non-setting mode, when run finishes, long press this button for 3 sec. to rerun; in setting mode, press this button to decrease set value, long press this button to decrease continuously.

5. “increase/backlight” button: in non-setting mode, press this button to turn on or off  the screen backlight; in setting mode, press this button to increase set value, long press this button to increase continuously.
 
6. "ALM!" indicator light: alarm light, there is an over-temperature alarm or temperature overflow alarm, the indicator light.

7. "HEAT" indicator light: heating indicator light, with heating output, the indicator light.

8. "AT" indicator: self-settlement indicator, when self-settlement timing, the indicator blinks.
9. "RUN" indicator light: This indicator light is lit when the instrument is operating normally.

Then start operating:

1. Start the equipment, screen display will light up. Wait for 2 seconds to enter the starter mode.

2. Set temp, time and fan rpm.

Press “Set” button, enter “temperature setting” mode, upper screen flashing, you can press “shift”, “increase”, “decrease” to reach set values needed; press “set” button again, enter “time setting” mode, two characters on the left hand side of lower screen flash, press “shift”, “increase”, “decrease” buttons to set “hour” value you need; then press “set” button, two characters on the right hand side of lower screen flash, press “shift”, “increase”, “decrease” to reach set “minute” value; press “set” button one more time, fan speed flashes, press “shift”, “increase”, “decrease” to reach set value you need (skip this step if fan is not installed); press “set” button again, exit set mode and auto-save values.

 When set value for time is “0”, means no timing function, controller will run continuously. Lower screen displays set value of temperature; when set value for time is not “0”, lower screen displays time and the “second” symbol flashes when test temperature reaches set temperature. When time is up, it will stop running and shut down all output parameters. Lower screen shows “End”, beeper rings 3 times. Once it stops, long press “decrease/rerun” button for 3 sec. to restart.

3. “ALM”alarm indicator is on when is over heating. Beeper beeps continuously. Press any button to mute.

4. In setting mode, no press any button within 60 seconds, controller returns back to starter mode automatically.

5. If upper screen shows “----”, it could be failure of temperature sensor or the controller. Please check temp. sensor and wiring connection.

The above is how to set the temperature of the drying oven. Kenton laboratory are happy to communicate with you!

Share to
  • Pre-sale
  • 0086-020-36246586
  • 0086-020-36247961
  • 0086-020-36246649
  • After-sales
  • 0086-18688422996
  • Complaint
  • service@kentonchina.com