The definition of the new version of the liquid crystal temperature gauge and vacuum gauge of the vacuum drying oven

Kenton laboratory has updated the vacuum drying oven KZ series control laboratory with new LCD temperature and vacuum gauges, what are the specific definitions? Here are the details for you.
\
LCD temperature gauge
\
(1) "Main interface" indicator: this lamp is lit during normal operation, and goes out when entering the setting state.
(2) "RUN" indicator: This lamp is off at the end of timer, otherwise it is always on. 3
(3) "STOP" light: This light turns on at the end of timer, otherwise it turns off.
(4) "AT" light: This light turns off at the end of timer, otherwise it turns off.
(5) "Heating" indicator light: this light is lit when heating output, otherwise it goes out.

 
Vacuum table
\

1. PV measurement value window: displaying the measured vacuum value in normal state and prompting characters in setting state.
2.SV set value window: display the set vacuum value in normal state and the parameter value in set state.
3.IN Inlet light: the light indicates that the inlet solenoid valve is open and the vacuum value increases.
4.OUT pumping light: the light indicates that the pumping solenoid valve is open and the vacuum value is reduced.
5. Pump light: the light indicates that the pumping vacuum pump is open, after the pumping vacuum pump is open, the pumping solenoid valve can effectively pump the vacuum and the vacuum degree value will be reduced.
6.ALM alarm light: the light indicates that the vacuum setting value cannot be reached for a long time and needs to be checked.
7.AIR Deflate light: the light indicates that the deflate solenoid valve is open and the air inside the oven is in balance with the outside world.
8. Left Setting key: under normal state, click to set SV, ST; press and hold for 3 seconds to enter setting state.
Press and hold for 3 seconds to exit the setting state.
9. STEP Shift key: click to view the current total number of segments and the number of segments running in normal conditions.
Switching the set point digit in the set state.
CYC Decrease key: To view the current total cycle and the operating cycle in normal state.
Decreasing the set point in the set state.
11.RES Increase key: press and hold for 3 seconds to restart or stop the operation when it is stopped.
Increasing the set point in the set state.
12. Right first defoliator: press and hold for 2 seconds to open the defoliator solenoid valve.
Press and hold for another 2 seconds to close the bleed solenoid valve.