Name
|
|
Initial value (setting range)
|
Timer
|
Year, month, day, hour, minute
When the current time at start-up is less than that time, the system enters the appointment waiting state until the current time is greater than or equal to that time, when the system enters the running state. When the current time at start-up is greater than or equal to that time, the system enters the running state. |
December 1, 2010 at 0:0 pm |
Timing methods
|
Start timing: the system can jump directly from the running state to the timing state.
Constant temperature timing: when the system is running, and when "main temperature set value - constant temperature timing deviation ≤ main temperature measured value ≤ main temperature set value + constant temperature timing deviation", the system can enter the timing state. |
Start Timer |
Constant temperature timing
deviation |
If the measured temperature value differs from the set value within this parameter, the system goes from "running" to "timing" and starts timing.
|
(0~10.0℃) 0.5 |
Alarm
|
A beep will sound at the end of the run to indicate the time.
Note: 0 means continuous beeping. |
(0~300s) 60 |
Lighting timer
|
Automatic switch-off time after lighting has been turned on. (System settings can be set with or without the lighting function)
Note: 0 means the lighting must be turned off manually. |
(0~9999min) 1 |