The environmental requirements use of Precision LCD screen electric heating incubator

When using Precision LCD screen electric heating incubator, users in the parameter table or manual will see "ambient temperature" and "temperature range" and so on, it can be said that the two are completely different concepts, "ambient temperature" and the "temperature range" has a direct impact. According to the industry standard YY0027-90 electric thermostat incubator on the use of the incubator environmental conditions have the following requirements. a) The temperature is 5~40℃ b) Relative humidity of not more than 90% c) Atmospheric pressure of 86kPa ~ 106kPa d) No strong vibration and corrosive gases around the incubator e) Avoid the influence of direct sunlight or other hot and cold sources For the ambient temperature, the general situation is called room temperature, in the room temperature 5 ~ 40 ℃ Under the premise that the incubator can be set within the temperature range parameters and constant temperature, that is, room temperature +5 ℃ -65 ℃ range. For example, in the use of room temperature 20 ℃, then the set temperature is at 25 ℃ -65 ℃, in the use of room temperature 30 ℃ set temperature is 35-65 ℃ range. But when the ambient temperature is lower than 5 ℃ or higher than 40 ℃ temperature range may deviate.